http://openehr.org/ckm/ Use to record measurements from a single acoustic reflext test, for diagnostic or screening purposes. Ipsilateral testing, contralateral testing or both, can be recorded per test ear. Use to record the interpretation of all measurements for each single ear test configuration, and an overall interpretation (or audiological diagnosis). Acoustic reflexes measure the eardrum movement generated by the stapedius and tensor tympani reflex in response to intense sound. The acoustic reflex threshold can be helpful in checking for particular types of hearing loss in situations where patient reliability is questionable. Acoustic reflex testing can also be done as part of a battery of tests to investigate central nervous system pathology, although this usage is now uncommon.

collegamento

http://openehr.org/ckm/

Keywords

  1. 10/04/18 10/04/18 - Halim Ugurlu
Titolare del copyright

openEHR Foundation

Caricato su

10 aprile 2018

DOI

Per favore, per richiedere un accesso.

Licenza

Creative Commons BY-NC 3.0

Commenti del modello :

Puoi commentare il modello dati qui. Tramite i fumetti nei gruppi di articoli e articoli è possibile aggiungere commenti a quelli in modo specifico.

Commenti del gruppo di articoli per :

Commenti dell'articolo per :

Per scaricare i modelli di dati devi essere registrato. Per favore accesso o registrati GRATIS.

openEHR-EHR-OBSERVATION.acoustic_reflex_result.v1

openEHR-EHR-OBSERVATION.acoustic_reflex_result.v1

  1. StudyEvent: openEHR-EHR-OBSERVATION.acoustic_reflex_result.v1
    1. openEHR-EHR-OBSERVATION.acoustic_reflex_result.v1
openEHR-EHR-OBSERVATION.acoustic_reflex_result.v1.xml
Descrizione

openEHR-EHR-OBSERVATION.acoustic_reflex_result.v1.xml

Alias
UMLS CUI-1
C0034934
Acoustic Reflex Test Result
Descrizione

Record of measurements from an acoustic reflex test, and their clinical interpretation.

Tipo di dati

text

Alias
UMLS CUI [1]
C0034934
Event Series
Descrizione

Event Series

Tipo di dati

text

Alias
UMLS CUI [1,1]
C0441471
UMLS CUI [1,2]
C0205549
Any Point-in-time Event
Descrizione

Default, unspecified point in time event which may be explicitly defined in a template or at run-time.

Tipo di dati

text

Alias
UMLS CUI [1]
C0439547
Test Result Name
Descrizione

Identification of the acoustic reflex test performed.

Tipo di dati

text

Alias
UMLS CUI [1,1]
C0456984
UMLS CUI [1,2]
C0027365
Result Details
Descrizione

The test measurements and interpretations which can be recorded per ear, and includes ipsilateral and contralateral testing.

Tipo di dati

text

Alias
UMLS CUI [1,1]
C1274040
UMLS CUI [1,2]
C1522508
Test Ear Configuration
Descrizione

Identification of the ear to which the stimulus is presented and the ear where the probe is situated.

Tipo di dati

text

Alias
UMLS CUI [1]
C0849529
Diagnostic Test
Descrizione

Reflex test that measures threshold levelss, reflex latency, reflex amplitude and reflex decay.

Tipo di dati

text

Alias
UMLS CUI [1]
C0456736
Screening Result
Descrizione

Reflex test that identifies the presence or absence of a reflex at a predetermined stimulation level.

Tipo di dati

text

Alias
UMLS CUI [1,1]
C0456736
UMLS CUI [1,2]
C1274040
Result
Descrizione

Result of the test.

Tipo di dati

text

Alias
UMLS CUI [1]
C1274040
No Screening Result
Descrizione

No screening test result is available for the test ear configuration.

Tipo di dati

boolean

Alias
UMLS CUI [1,1]
C1274040
UMLS CUI [1,2]
C1298908
Reason For No Test Result
Descrizione

Reason why no screening test result is available for the test ear configuration.

Tipo di dati

text

Alias
UMLS CUI [1,1]
C0566251
UMLS CUI [1,2]
C3841220
Threshold Level
Descrizione

The lowest stimulation level at which a reflex could be measured by the probe.

Tipo di dati

float

Alias
UMLS CUI [1]
C0449864
Clinical Interpretation
Descrizione

Clinical interpretation of all diagnostic results for the specified test ear configuration.

Tipo di dati

text

Alias
UMLS CUI [1]
C4086187
Test Signal
Descrizione

Identification of the stimulation signal used for the test.

Tipo di dati

text

Alias
UMLS CUI [1,1]
C1710082
UMLS CUI [1,2]
C0018786
Reflex Latency
Descrizione

Length of time from onset of the stimulation tone to the onset of the middle ear muscle reflex.

Tipo di dati

float

Alias
UMLS CUI [1,1]
C0034934
UMLS CUI [1,2]
C0457404
Reflex Decay
Descrizione

Length of time that the reflex amplitude can be sustained.

Tipo di dati

float

Alias
UMLS CUI [1]
C0846632
No Diagnostic Result
Descrizione

No diagnostic test result is available for the test ear configuration.

Tipo di dati

boolean

Alias
UMLS CUI [1,1]
C0456984
UMLS CUI [1,2]
C1298908
Reason For No Test Result
Descrizione

Reason why no diagnostic test result is available for the test ear configuration.

Tipo di dati

text

Alias
UMLS CUI [1,1]
C0566251
UMLS CUI [1,2]
C3841220
Overall Interpretation
Descrizione

Overall clinical interpretation of all of the results.

Tipo di dati

text

Alias
UMLS CUI [1]
C0420833
Intensity
Descrizione

Clinical interpretation of reflex threshold during diagnostic testing, based on intensity.

Tipo di dati

text

Alias
UMLS CUI [1,1]
C4086187
UMLS CUI [1,2]
C0449864
UMLS CUI [1,3]
C0034934
UMLS CUI [1,4]
C0522510
Result
Descrizione

The result of the test.

Tipo di dati

text

Alias
UMLS CUI [1]
C1274040
Comment
Descrizione

Additional narrative about the test results and intepretation not captured in other fields.

Tipo di dati

text

Alias
UMLS CUI [1]
C0947611
Test Result Image
Descrizione

Digital representation of the entire result.

Tipo di dati

text

Alias
UMLS CUI [1,1]
C0920316
UMLS CUI [1,2]
C0376478
Sensation Level
Descrizione

Clinical interpretation of reflex Threshold Level result based on sensation level.

Tipo di dati

text

Alias
UMLS CUI [1]
C0036658
Clinical Interpretation
Descrizione

Clinical interpretation of the screening test result.

Tipo di dati

text

Alias
UMLS CUI [1]
C4086187
Calibration Reference dB
Descrizione

Scale used for acoustic calibration of the test signal.

Tipo di dati

text

Alias
UMLS CUI [1,1]
C0006751
UMLS CUI [1,2]
C0439497
Criteria for Screening Assessment Pass
Descrizione

The criteria by which the Screening Assessment is passed.

Tipo di dati

text

Alias
UMLS CUI [1,1]
C1545319
UMLS CUI [1,2]
C0243161
UMLS CUI [1,3]
C2828386
Intensity
Descrizione

The loudness of the screening stimulus.

Tipo di dati

float

Alias
UMLS CUI [1,1]
C0178733
UMLS CUI [1,2]
C0028263
Frequency
Descrizione

The frequency of the test signal.

Tipo di dati

float

Alias
UMLS CUI [1]
C0439603
Test Signal Type
Descrizione

The type of stimulation used to elicit a reflex.

Tipo di dati

text

Alias
UMLS CUI [1,1]
C0521310
UMLS CUI [1,2]
C0871511
Immittance Machine
Descrizione

Details of immittance machine used to conduct the test.

Tipo di dati

text

Alias
UMLS CUI [1,1]
C0336779
UMLS CUI [1,2]
C2724352
Confounding Factors
Descrizione

Description of incidental factors that may be contributing to the test result.

Tipo di dati

text

Alias
UMLS CUI [1]
C0009673
Comment
Descrizione

Additional narrative about the method of testing not captured in other fields.

Tipo di dati

text

Alias
UMLS CUI [1]
C0947611
Probe Tone Frequency
Descrizione

The frequency of the probe tone.

Tipo di dati

text

Alias
UMLS CUI [1,1]
C0237917
UMLS CUI [1,2]
C0182400
Normal Range Definition
Descrizione

Definition of the 'Normal Range' value specified within the Intensity data element used for Clinical Interpretation of the Diagnostic Test.

Tipo di dati

text

Alias
UMLS CUI [1,1]
C1704788
UMLS CUI [1,2]
C0086715

Similar models

openEHR-EHR-OBSERVATION.acoustic_reflex_result.v1

  1. StudyEvent: openEHR-EHR-OBSERVATION.acoustic_reflex_result.v1
    1. openEHR-EHR-OBSERVATION.acoustic_reflex_result.v1
Name
genere
Description | Question | Decode (Coded Value)
Tipo di dati
Alias
Item Group
openEHR-EHR-OBSERVATION.acoustic_reflex_result.v1.xml
C0034934 (UMLS CUI-1)
Acoustic Reflex Test Result
Item
Acoustic Reflex Test Result
text
C0034934 (UMLS CUI [1])
Event Series
Item
Event Series
text
C0441471 (UMLS CUI [1,1])
C0205549 (UMLS CUI [1,2])
Any Point-in-time Event
Item
Any Point-in-time Event
text
C0439547 (UMLS CUI [1])
Item
Test Result Name
text
C0456984 (UMLS CUI [1,1])
C0027365 (UMLS CUI [1,2])
Code List
Test Result Name
CL Item
Acoustic Reflex Test (1)
CL Item
Non Acoustic Reflex Test (2)
Result Details
Item
Result Details
text
C1274040 (UMLS CUI [1,1])
C1522508 (UMLS CUI [1,2])
Item
Test Ear Configuration
text
C0849529 (UMLS CUI [1])
Code List
Test Ear Configuration
CL Item
Left Ipsilateral (1)
CL Item
Right Ipsilateral (2)
CL Item
Left Contralateral (3)
CL Item
Right Contralateral (4)
Diagnostic Test
Item
Diagnostic Test
text
C0456736 (UMLS CUI [1])
Screening Result
Item
Screening Result
text
C0456736 (UMLS CUI [1,1])
C1274040 (UMLS CUI [1,2])
Item
Result
text
C1274040 (UMLS CUI [1])
Code List
Result
CL Item
Present (1)
CL Item
Absent (2)
No Screening Result
Item
No Screening Result
boolean
C1274040 (UMLS CUI [1,1])
C1298908 (UMLS CUI [1,2])
Reason For No Test Result
Item
Reason For No Test Result
text
C0566251 (UMLS CUI [1,1])
C3841220 (UMLS CUI [1,2])
Threshold Level
Item
Threshold Level
float
C0449864 (UMLS CUI [1])
Clinical Interpretation
Item
Clinical Interpretation
text
C4086187 (UMLS CUI [1])
Item
Test Signal
text
C1710082 (UMLS CUI [1,1])
C0018786 (UMLS CUI [1,2])
Code List
Test Signal
CL Item
Pure Tone - 500 Hz (1)
CL Item
Pure Tone - 1000 Hz (2)
CL Item
Pure Tone - 2000Hz (3)
CL Item
Pure Tone - 4000Hz (4)
CL Item
Broad Band Noise (5)
CL Item
Narrow Band Noise (6)
Reflex Latency
Item
Reflex Latency
float
C0034934 (UMLS CUI [1,1])
C0457404 (UMLS CUI [1,2])
Reflex Decay
Item
Reflex Decay
float
C0846632 (UMLS CUI [1])
No Diagnostic Result
Item
No Diagnostic Result
boolean
C0456984 (UMLS CUI [1,1])
C1298908 (UMLS CUI [1,2])
Reason For No Test Result
Item
Reason For No Test Result
text
C0566251 (UMLS CUI [1,1])
C3841220 (UMLS CUI [1,2])
Overall Interpretation
Item
Overall Interpretation
text
C0420833 (UMLS CUI [1])
Item
Intensity
text
C4086187 (UMLS CUI [1,1])
C0449864 (UMLS CUI [1,2])
C0034934 (UMLS CUI [1,3])
C0522510 (UMLS CUI [1,4])
Code List
Intensity
CL Item
Normal Range (1)
CL Item
Absent (2)
CL Item
Elevated (3)
Result
Item
Result
text
C1274040 (UMLS CUI [1])
Comment
Item
Comment
text
C0947611 (UMLS CUI [1])
Test Result Image
Item
Test Result Image
text
C0920316 (UMLS CUI [1,1])
C0376478 (UMLS CUI [1,2])
Item
Sensation Level
text
C0036658 (UMLS CUI [1])
Code List
Sensation Level
CL Item
Normal (1)
CL Item
Reduced (2)
CL Item
Elevated (3)
CL Item
Absent (4)
Clinical Interpretation
Item
Clinical Interpretation
text
C4086187 (UMLS CUI [1])
Item
Calibration Reference dB
text
C0006751 (UMLS CUI [1,1])
C0439497 (UMLS CUI [1,2])
Code List
Calibration Reference dB
CL Item
dB SPL (1)
CL Item
dB HL (2)
Criteria for Screening Assessment Pass
Item
Criteria for Screening Assessment Pass
text
C1545319 (UMLS CUI [1,1])
C0243161 (UMLS CUI [1,2])
C2828386 (UMLS CUI [1,3])
Intensity
Item
Intensity
float
C0178733 (UMLS CUI [1,1])
C0028263 (UMLS CUI [1,2])
Frequency
Item
Frequency
float
C0439603 (UMLS CUI [1])
Test Signal Type
Item
Test Signal Type
text
C0521310 (UMLS CUI [1,1])
C0871511 (UMLS CUI [1,2])
Immittance Machine
Item
Immittance Machine
text
C0336779 (UMLS CUI [1,1])
C2724352 (UMLS CUI [1,2])
Confounding Factors
Item
Confounding Factors
text
C0009673 (UMLS CUI [1])
Comment
Item
Comment
text
C0947611 (UMLS CUI [1])
Item
Probe Tone Frequency
text
C0237917 (UMLS CUI [1,1])
C0182400 (UMLS CUI [1,2])
Code List
Probe Tone Frequency
CL Item
226Hz (1)
CL Item
1000Hz (2)
Normal Range Definition
Item
Normal Range Definition
text
C1704788 (UMLS CUI [1,1])
C0086715 (UMLS CUI [1,2])

Si prega di utilizzare questo modulo per feedback, domande e suggerimenti per miglioramenti.

I campi contrassegnati da * sono obbligatori.

Do you need help on how to use the search function? Please watch the corresponding tutorial video for more details and learn how to use the search function most efficiently.

Watch Tutorial